Skip Ribbon Commands
Skip to main content

Equipment Detail

Name

CFX96 Touch™ Real-Time PCR Detection System

Manufacturer

BioRad

Make or Model

n/a

Description

​The CFX96™ Optical Reaction Module converts the C1000 Touch™ Thermal Cycler chassis into the powerful and precise CFX96 Touch Real-Time PCR Detection System. This six-channel system combines advanced optical technology with precise thermal control to deliver sensitive, reliable detection. Quickly set up runs and monitor amplification traces in real time on the integrated LCD touch screen. With up to five-target detection, unsurpassed thermal cycler performance, unrivaled stand-alone functionality, and powerful yet easy-to-use software, the CFX96 Touch System is designed to advance your qPCR.

Thermal Cycler  
Chassis C1000 Touch™
Maximum ramp rate, °C/sec 5
Average ramp rate, °C/sec 3.3
Heating and cooling method Peltier
Lid, °C Heats up to 105
Temperature  
Range, °C 0–100
Accuracy, °C ±0.2 of programmed target at 90°C
Uniformity, °C ±0.4 well-to-well within 10 sec of arrival at 90°C
Gradient  
Operational range, °C 30–100
Programmable span, °C 1–24
Optical Detection  
Excitation 6 filtered LEDs
Detection 6 filtered photodiodes
Range of excitation/emission wavelengths, nm 450–730
Sensitivity Detects 1 copy of target sequence in human genomic DNA
Dynamic range 10 orders of magnitude
Scan Time  
All channels, sec 12
FAM/SYBR® Green only, sec 3
Software  
Operating systems Windows 7, Windows 8
Multiplex analysis Up to 5 targets per well
System  
Licensed for real-time PCR Yes
Sample capacity, wells 96
Sample size, µl 1–50 (10–25 recommended)
Communication interface USB 2.0
Electrical approvals IEC, CE
Dimensions (W x D x H), cm/in 33 x 46 x 36 /13 x 18 x 14
Weight, kg/lb 21/47

 

External Link

http://www.bio-rad.com/en-us/product/cfx96-touch-real-time-pcr-detection-system